285
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide
26 I2C Triggering and Serial
Decode
Setup for I2C Signals / 285
I2C Triggering / 286
I2C Serial Decode / 290
I2C triggering and serial decode is available on DSOX1200-Series and
EDUX1052A/G oscilloscopes.
Setup for I2C Signals
I
2
C (Inter-IC bus) signals setup consists of connecting the oscilloscope to the
serial data (SDA) line and the serial clock (SCL) line and then specifying the input
signal threshold voltage levels.
To set up the oscilloscope to capture I
2
C signals, use the
Signals
softkey which
appears in the Analyze menu:
1
Press
[Display] > Labels > Display
to turn on labels.
2
Press the
[Analyze]
key.
3
Press
Features
; then, select
Serial Bus
.
4
Press
Features
again (or push the Entry knob) to enable the serial bus display.
5
Press the
Mode
softkey; then, select
I2C
.