294
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
27
LIN Triggering and Serial Decode
6
Press the
Signals
softkey to open the LIN Signals menu.
7
Press the
Source
softkey to select the channel connected to
the LIN signal line.
The label for the LIN source channel is automatically set.
8
Press the
Threshold
softkey; then, turn the Entry knob to set
the LIN signal threshold voltage level to the middle of the LIN
signal.
The threshold voltage level is used in decoding, and it will
become the trigger level when the trigger type is set to the
selected serial decode slot.
9
Press the
Baud Rate
softkey to open the LIN Baud Rate menu.
10
Press the
Baud
softkey; then, turn the Entry knob to select the
baud rate that matches match your LIN bus signal.
The default baud rate is 19.2 kb/s.
If none of the pre-defined selections match your LIN bus
signal, select
User Defined
; then, press the
User Baud
softkey and turn the Entry
knob to enter the baud rate.
You can set the LIN baud rate from 2.4 kb/s to 625 kb/s in increments of
100 b/s.
11
Press the
Back key to return to the LIN Signals menu.
12
Press the
Sample Point
softkey; then, turn the Entry knob to select the sample
point at which the oscilloscope will sample the bit value.
13
Press the
Standard
softkey; then, turn the Entry knob to select the LIN standard
you are measuring (LIN 1.3 or LIN 2.0).
Back
60%
70%
80%
Sample Point
One Bit