81
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide
8 Serial Bus Decode/Trigger
Depending on the oscilloscope model, the following hardware-accelerated serial
decode and trigger features are available:
Triggering on
Serial Data
In some cases, such as when triggering on a slow serial signal (for example, I2C,
SPI, CAN, LIN, etc.), to stabilize the waveform display, it may be necessary to
switch from the Auto trigger mode to the Normal trigger mode to prevent the
oscilloscope from auto-triggering. You can select the trigger mode by pressing the
[Trigger]
key, then the
Mode
softkey.
Also, the threshold voltage level must be set appropriately for each source
channel. The threshold level for each serial signal can be set in the Signals menu
(press
[Analyze] > Features
, select
Serial Bus
, then press the
Signals
softkey).
See Also
•
, “CAN Triggering and Serial Decode,” starting on page 275
•
, “I2C Triggering and Serial Decode,” starting on page 285
•
, “LIN Triggering and Serial Decode,” starting on page 293
Serial decode and trigger type:
Available on:
CAN (Controller Area Network)
DSOX1200-Series models
I2C (Inter-IC)
DSOX1200-Series and EDUX1052A/G
models
LIN (Local Interconnect Network)
DSOX1200-Series models
SPI (Serial Peripheral Interface)
DSOX1200-Series models
UART (Universal Asynchronous Receiver/Transmitter)
protocols including RS232 (Recommended Standard 232)
DSOX1200-Series and EDUX1052A/G
models