148
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
14
Cursors
•
Hex
— Logic levels of displayed waveforms at the current X1 and X2 cursor
positions are displayed in hexadecimal.
Manual
and
Track Waveform
modes can be used on
waveforms that are displayed on the analog input
channels (including math functions).
In
Hex
and
Binary
modes, a level can be displayed as 1
(higher than trigger level), 0 (lower than trigger level),
indeterminate state ( ), or X (don't care).
In
Binary
mode, X is displayed if the channel is turned off.
In
Hex
mode, the channel is interpreted as 0 if turned off.
4
Press
Source
(or
X1 Source
,
X2 Source
in the
Track Waveform
mode); then, select the input source for cursor values.
5
Select the cursor(s) to be adjusted:
•
Push the Cursors knob; then, turn the Cursors knob. To
finalize your selection, either push the Cursors knob
again or wait about five seconds for the popup menu to
disappear.
Or:
•
Press the
Cursors
softkey; then, turn the Entry knob.
The
X1 X2 linked
and
Y1 Y2 linked
selections let you adjust both cursors at the
same time, while the delta value remains the same. This can be useful, for
example, for checking pulse width variations in a pulse train.
The currently selected cursor(s) display brighter than the other cursors.
6
To change the cursor units, press the
Units
softkey.