Mask Testing
16
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
181
9
To clear the mask and switch off mask testing, press the
Back key to return
to the Analyze menu, then press the
Clear Mask
softkey.
If infinite persistence display mode (see
is "on" when mask test is enabled, it stays on. If infinite persistence is "off" when
mask test is enabled, it is switched on when mask test is switched on, then infinite
persistence is switched off when mask test is switched off.
Troubleshooting
Mask Setup
If you press
Create Mask
and the mask appears to cover the entire screen, check the
± Y and ± X settings in the Automask menu. If these are set to zero the resulting
mask will be extremely tight around the waveform.
If you press
Create Mask
and it appears that no mask was created, check the ± Y
and ± X settings. They may be set so large that the mask is not visible.
Back