298
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
27
LIN Triggering and Serial Decode
4
Press the
Mode
softkey; then, select
LIN
.
5
Choose whether to include the parity bits in the identifier
field.
a
If you want to mask the upper two parity bits, ensure that
the box under the
Show Parity
softkey is not selected.
b
If you want to include the parity bits in the identifier field,
ensure that the box under the
Show Parity
softkey is
selected.
6
If the oscilloscope is stopped, press the
[Run/Stop]
key to
acquire and decode data.
You can use the horizontal
Zoom
window for easier navigation of the decoded data.
See Also
•
•
"Interpreting LIN Lister Data"
NOTE
If the setup does not produce a stable trigger, the LIN signal may be slow enough that the
oscilloscope is auto-triggering. Press the
[Trigger]
key, then press the
Mode
softkey to set the
trigger mode from
Auto
to
Normal
.