Measurements
15
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
167
Time Measurements
The following figure shows time measurement points.
The default lower, middle, and upper measurement thresholds are 10%, 50%, and
90% between Top and Base values. See
for other percentage threshold and absolute value threshold settings.
•
•
•
•
•
•
•
•
•
•
•
•
•
Period
Period is the time period of the complete waveform cycle. The time is measured
between the middle threshold points of two consecutive, like-polarity edges. A
middle threshold crossing must also travel through the lower and upper threshold
levels which eliminates runt pulses. The X cursors show what portion of the
waveform is being measured. The Y cursor shows the middle threshold point.
Rise Time
+ Width
Fall Time
- Width
Period
Middle
Upper
Thresholds
Lower