Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
7
Reference Waveforms
2
Press the
Skew
softkey and turn the Entry knob to adjust the reference
waveform skew.
To display reference waveform information
1
Press the
[Analyze]
key.
2
Press
Features
; then, select
R1
or
R2
.
3
Press
Save/Clear
.
4
In the Save/Clear menu, press the
Display Info
softkey to enable or disable
reference waveform information on the oscilloscope display.
5
Press the
Transparent
softkey to enable or disable transparent information
backgrounds.
This setting is also used for other oscilloscope information on the display, like
mask test statistics, etc.
To save/recall reference waveform files to/from a USB storage
device
Analog channel, math, or reference waveforms can be saved to a reference
waveform file on a USB storage device. See
"To save reference waveform files to a
You can recall a reference waveform file from a USB storage device into one of the
reference waveform locations. See