224
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
20
Save/Recall (Setups, Screens, Data)
To recall reference waveform files from a USB storage device
1
Press the
[Save/Recall]
key.
2
In the Save/Recall menu, press the
Recall
softkey.
3
In the Recall menu, press the
Recall
softkey and turn the Entry knob to select
Reference Waveform data (*.h5)
.
4
Press the
To Ref:
softkey and turn the Entry knob to select the desired reference
waveform location.
5
Press the softkey in the second position and use the Entry knob to navigate to
the file to recall. See
"To navigate storage locations"
6
Press the
Press to Recall
softkey.
A message indicating whether the recall was successful is displayed.
7
If you would like to clear the display of everything except the reference
waveform, press
Clear Display
.
Recalling Default Setups
1
Press the
[Save/Recall]
key.
2
In the Save/Recall menu, press
Default/Erase
.
3
In the Default menu, press one of these softkeys:
•
Default Setup
— recalls the oscilloscope's default setup. This is the same as
pressing the front panel
[Default Setup]
key. See
Some user settings are not changed when recalling the default setup.
•
Factory Default
— recalls the oscilloscope's factory default settings.
You must confirm the recall because there are no user settings that are left
unchanged.
Performing a Secure Erase
1
Press the
[Save/Recall]
key.
2
In the Save/Recall menu, press
Default/Erase
.