6 Electrical connection
Use shrink tubing to insulate the shield and Shield-Kon® connector
1
6.2.2
Conductor cross-sections for base unit and power supply unit
The terminals on the base unit and power supply unit are plug-in screw termi-
nals.
^
1.Shield-Kon is a registered trademark of THOMAS & BETTS INTERNATIONAL, Inc.,
Wilmington Del., US.
60 mm
Shield-Kon® connector
Ferrule
Conductor cross sec-
tion
Length to
strip
minimal
maximal
without ferrule
Power supply unit voltage supply 0.2 mm
²
2.5 mm²
7 mm
Power supply unit relay switching
outputs
0.2 mm
²
2.5 mm²
7 mm
Base unit
0.14 mm
2
1.5 mm
2
7 mm
Ferrule without lip
Power supply unit voltage supply 0.25 mm
2
2.5 mm²
7 mm
Power supply unit relay switching
outputs
0.25 mm
2
2.5 mm²
7 mm
Base unit
0.25 mm
2
1.5 mm
2
7 mm
with ferrule with lip
Power supply unit voltage supply 0.25 mm
2
2.5 mm²
7 mm
Power supply unit relay switching
outputs
0.25 mm
2
2.5 mm²
7 mm
Base unit
0.25 mm
2
0.5 mm
2
7 mm
Rigid
Power supply unit voltage supply 0.2 mm
²
2.5 mm²
7 mm
Power supply unit relay switching
outputs
0.2 mm
²
2.5 mm²
7 mm
Base unit
0.14 mm
2
1.5 mm
2
7 mm
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......