6 Electrical connection
IN 7
IN 8
pH/Redox
Symmetric connection of a combination electrode with
integrated resistance thermometer and VarioPin plug
head
Symmetric connection is used to reduce interference from
stray electromagnetic fields along the sensor cable.
The RTD temperature probe is used to provide a tempera-
ture-compensated pH-value measurement, and can be con-
nected to a temperature input or universal input.
A = Glass/metal electrode (core)
B = Reference electrode (inner shield)
C = Liquid potential (grounding pin, pipe, or container wall at
the measuring point)
D = Shield (outer shield)
E = RTD temperature probe
F = RTD temperature probe
G = RTD temperature probe
Terminal 2 is not connected!
Slot
Option/connection variant
Symbol
?
A
B
E
F
G
D
C
1
3
5
6
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......