17 Calibrating Ci conductivity sensors
200
Calibration default settings for calibrating the TC curve
Parameter
Setting options
Explanation
Temperature com-
pensation
Selection from
analog selection
Temperature input for automatic sensing of the test/
sample solution temperature during the calibration
Starting temperature
-50 to 250 °C
The starting and final temperatures of the range for
which a temperature coefficient curve is to be deter-
mined.
The starting temperature must be at least 20 °C low-
er than the final temperature. The reference tem-
perature for the measuring input must lie between
the starting and final temperatures, and differ from
the starting and final temperatures by at least 2 °C.
Final temperature
-50 to 250 °C
NOTE!
Calibration of the TC curve is possible only with automatic temperature sens-
ing.
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......