10 Configuration
Text sensor alarm
up to 21 text characters
Text for the alarm/event list in case of sensor er-
ror
Alarms 1/2
Analog input alarms are used to monitor measured values
in relation to adjustable limit values.
The alarm settings for all analog device functions are explained together.
v
Chapter 10.9.2 "Alarms for analog signals", Page 135
Calibration timers
Calibration timers prompt the user to calibrate sensors on a regular basis.
The settings for all analysis inputs and the universal inputs are explained to-
gether.
v
Chapter 10.10 "Calibration timers", Page 139
a
The entry field units depend on the configuration items "Electrode type" and
"Redox unit".
NOTE!
The following points must be observed for correct operation of glass electrode
monitoring via impedance measurement (see preceding table):
•
Impedance measurements are possible only with glass-based sensors.
•
Sensors must be connected directly to an analysis input for pH/Redox/NH
3
on the device.
•
Impedance converters must not be installed in the measuring circuit.
•
The maximum admissible cable length between sensor and device is 10 m.
•
Fluid resistances have a direct impact on the measurement result. It is
therefore advisable to activate the impedance measurement in liquids at a
minimum conductivity of approx. 100 µS/cm.
Configuration point Selection/
setting option
Explanation
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......