13 Calibrating
a pH measuring chain
13.1
Notes
13.2
General information
The calibration of pH electrodes is based on measurements in Buffer solutions
with a defined pH-value. The pH values of the Buffer solutions used are speci-
fied either via entry of fixed values into the calibration default settings, entered
during the calibration or recognized automatically by "automatic buffer recogni-
tion" during the calibration process. For "automatic buffer recognition", a Buffer
set table must be selected in the calibration default settings. In this case, the
Buffer solutions used must be listed in the provided Buffer set table. Since pH
value measurement of liquids is temperature-dependent, the temperature of the
Buffer solution must be sensed in order to compensate for its effect on the result
of the measurement. This requires either manual entry or measurement with the
aid of a Temperature sensor.
13.2.1
Calibration methods for pH sensors
Zero-point calibration
This calibration method is used to determine the pH zero point on the measure-
ment characteristic curve. The Slope is retained.
A Buffer solution with a defined pH-value is needed as a reference.
Two-point Calibration
By measuring 2 different Buffer solutions with defined pH-values, the pH zero
point and pH slope of the measuring chain are established.
The pH-values of the Buffer solutions must be at least 2 pH apart. This calibra-
tion method is recommended for most applications.
Three-point calibration
With Three-point calibration, the pH zero point as well as the pH slope in the
acidic region and the pH slope in the alkaline region are established.
This method requires 3 Buffer solutions with defined pH-values as references.
One of these must be acidic, one neutral, and one alkaline. The pH values of the
Buffer solution must be at least 2 pH apart from one another. This calibration
method is recommended for applications with more demanding accuracy re-
quirements when performing measurements in both the alkaline and acidic re-
gions.
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......