![JUMO AQUIS touch P 202580 Operating Manual Download Page 315](http://html1.mh-extra.com/html/jumo/aquis-touch-p-202580/aquis-touch-p-202580_operating-manual_2031168315.webp)
315
20 Technical data
20.10 Electrical data
20.11 Screen Touchscreen
Voltage supply (switch-mode)
AC 110 to 240 V +10/-15%; 48 to 63 Hz or
AC/DC 24 V +30/-25 %; 48 to 63 Hz
electrical safety
According to EN 61010, part 1
overvoltage category III,
pollution degree 2
Max. power consumption
AC 110 to 240 V
27.9 VA (11,7 W at 230 V)
AC/DC 24 V
25.2 VA (14,7 W at DC 24 V)
Data backup
Flash memory
Electrical connection
at the back via screw terminals
Specifications for conductor cross-sections
v
Chapter 6.2.2 "Conductor cross-sections for base unit and
electromagnetic
compatibility (EMC):
DIN EN 61326-1
Interference emission
Class A
Interference immunity
Industrial requirements
Type
TFT-touchscreen
Touchscreen sensors
Resistive (can also be operated wearing gloves)
Display protection
Front/design film for protection against damage and scratches
Size
3.5"
Resolution
320 × 240 pixel
Color depth
256 colors
Viewing angle
Horizontal:
±
65°
Vertical: -65 to +40°
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......