6 Electrical connection
36
Use shrink tubing to insulate the braided shield
Preparing a shielded multiconductor cable
Strip the insulation off the connecting cable and pull back the shield
Attach wire ferrules to the ends of the conductors and a Shield-Kon® connector
1
for shielding
Inner conductor
Shrink sleeve
NOTE!
The black, semiconducting layer must not touch with the inside conductor! This
would short-circuit the signal from the pH electrode.
1.Shield-Kon is a registered trademark of THOMAS & BETTS INTERNATIONAL, Inc.,
Wilmington Del., US.
60 mm
60 mm
Shield-Kon® connector
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......