10 Configuration
158
K-Faktor
0 to. 99999 l
-1
Ratio of the pulse count to flow rate (pulses per
liter)
The K-factor can be obtained from the documen-
tation for the fitting in which the flow-through sen-
sor (e.g. paddle wheel sensor) is installed.
If the flow-through signal is transmitted as a stan-
dard signal
or via an interface (external analog inputs),
be aware that the K-factor affects analog values
as well. For reliable handling of the flow-through
as an analog signal, it is recommended that the
flow value be generated as a standard signal and
the K-Faktor set to a value of 1.
Unit
l / s
l / min
l / h
m
3
/ s
m
3
/ min
m
3
/ h
gal / s
gal / min
gal / h
customer-specific
Unit for the flow-through value
Customer-specific unit:
In the individual measured display, the
flow rate value in liters per second is multiplied by
the factor from the "Conversion factor" setting
and the unit displayed on the basis of the " Unit
description" setting.
Conversion factor
(flow rate)
-99999 to +99999
only for "customer-specific" as "Unit":
Conversion factor from the unit "l / s" for the
flow rate to the customer-specific unit
Unit description
(flow rate)
up to 5 text characters
only for "customer-specific" as "Unit":
Entry of an arbitrary customer-specific unit for the
flow rate
Decimal place
(flow rate)
Auto,
Fixed comma format
Decimal points in the display
Start
Display range
(flow rate)
-99999 to +99999
a
Upper/lower limit for labeling the scale
when displaying measuring values, e.g. in re-
corder diagrams and bar graphs
End
Display range
(flow rate)
-99999 to +99999
a
Configuration point Selection/
setting option
Explanation
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......