19 PC Setup Program
246
Setup dialog - USB interface settings on the PC
If several devices are connected via a USB interface, one device must be se-
lected here. Furthermore, various test criteria can be activated so that the con-
nection to the device is established only after a successful test.
This dialog depends on the selected connection preset. The figure shows the
case "Connection via USB". To access this dialog, it is necessary to click on the
"Next" button twice – starting from the "Device Login" dialog – while holding
down the "Ctrl" + "Shift" keys.
Adjustments
Clicking on the "Finish"
button closes the assistant and enters the new connec-
tion in the device connections list.
Setup dialog - Overview of the settings
Setting item
Selection/
setting option
Explanation
Connected de-
vices
Selection from
dropdown list
Devices connected via the USB inter-
face are displayed.
Name
Text
(entry field active
only when check-
box activated)
With checking active, the device is con-
nected only if the device name matches.
Ser. No.
With checking active, the device is con-
nected only if the device serial number
matches.
CPU
With checking active, the device is con-
nected only if the serial number of the
CPU matches.
Path
With checking active, the device is con-
nected only if the USB path matches.
Extended
Open by
pressing the button
Extended settings for communication
These settings should be changed only
in exceptional situations.
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......