10 Configuration
130
CR/Ci measuring range configuration
Open:
Device menu
r
Configuration
r
Analog inputs
r
Analysis input 1 to 4
r
Measuring range 1 to 4
Configuration point Selection/
setting option
Explanation
TDS factor
0.01 to 2.00
only for conductive conductivity with
TDS compensation:
Conversion factor from measured conductivity to
display unit (see configuration item "Unit" in this
table)
Compensation
for CR/Ci:
Off,
TC linear,
natural water,
natural water with
expanded temperature
range
Type of temperature compensation for conduc-
tivity measurement
only for CR:
TDS,
ASTM neutral,
ASTM acid,
ASTM alkaline
only for Ci:
TC curve,
NaOH 0 to 12%,
NaOH 25 to 50%,
HNO3 0 to 25%,
HNO3 36 to 82%,
H2SO4 0 to 28%,
H2SO4 36 to 85%,
H2SO4 92 to 99%,
HCL 0 to 18%,
HCL 22 to 44%
Unit for
calculation
for CR/Ci:
µS/cm
mS/cm
Unit in which the conductivity is displayed
only for CR:
k
Ω
/cm
M
Ω
/cm
Unit
up to 5 text characters
only for conductive conductivity measure-
ment with
TDS compensation:
Unit for the process variable to be displayed for
TDS measurements or when using customer-
specific linearization (e.g. ppm or mg/l)
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......