123
10 Configuration
10.5.2
Universal inputs of base unit and optional boards
Base unit universal input: IN 6
Optional board universal inputs: IN 11/12
Open:
Device menu
r
Configuration
r
Analog inputs
r
Universal input 1 to 3
r
Configuration
Configuration point Selection/
setting option
Explanation
Description
up to 15 text characters
Designation for the input
Operating mode
linear scaling,
temp. measurement,
pH-value measurement,
conductivity measure-
ment,
free chlorine
pH/T-compensated
Type of measuring
linear scaling:
Standard signals (for
IN 11/12 and resistance potentiometer
a
) with lin-
ear measurement characteristic line or customer-
specific linearization
For standard signals the scale start point, scale
end point and unit must be specified.
Temp. measurement:
Measurement made with
a
resistance thermometer
The type of sensor is selected in the "Signal type"
configuration item. The unit for the temperature is
specified in the "Basic Settings" menu.
v
Chapter 10.2 "Basic settings", Page 116
pH value, conductivity, and free chlorine:
The
values measured by the individual analysis sen-
sors are received in the form of a standard signal.
Outside factors affecting the individual analysis
process variables are compensated. It is thus
necessary to make appropriate compensation
settings in the configuration of the universal in-
put.
Signal type
IN 6/11/12:
0 to 20 mA
4 to 20 mA
20 to 0 mA
20 to 4 mA
Type of sensor connected
Correct scale settings are required for the
stan-
dard signals
v
"Scale start/end" in this table.
appropriate linearizations are available for
Pt100, Pt1000 and
resistance potentiometer
a
A customer -specific linearization must be config-
ured for
400
Ω
and 4000
Ω
v
"Customer-specific linearization" in this table
only IN 11/12:
0 to 10 V
10 to 0 V
Pt100
Pt1000
400
Ω
4000
Ω
Resistance potentiome-
ter
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......