11 Retrofitting optional boards
166
4
Place the wire of the calibration sensor with 2 windings through the
opening in the Ci sensor without connecting the ends of the wires.
5
Start the Ci base calibration
Device Menu
r
Service
r
Ci Base Calibration IN 7 to 8
6
Enter the cell constant for the sensor and confirm by pressing "OK".
7
The device now performs a measuring with the open cable loop of
the calibration adapter open. Wait until the measured value dis-
played stabilizes and then confirm by pressing "OK".
Step
Action
Tap button
to enter