Analog inputs
Proper operation of the analog inputs of the device can be checked here.
In addition to selecting the analog input and signal type, the line frequency of the
power supplied to the device must be set in order to test an analog input correct-
ly. Pressing the "Test" button displays a connection diagram for the test signal.
Ensure that a valid test signal acc. to the connection diagram is present at the
input. Confirming the connection diagram starts the test. The measuring values
coming from the analog input concerned can be viewed in the actual-value dis-
play and in the measuring value table. To stop the test, press the "Stop" button.
Open:
Online Parameters
r
Adjusting / Testing (double-click)
r
Analog inputs tab
(1)
"Stop" button
Stops a test
(2)
"Test" button
Starts a test
(3)
Actual-value display
Displays the actual analog value at the input
(4)
Measuring value table
Lists the measuring values over the course of time
(5)
"Supply frequency" setting
(6)
Selection of the analog input to test
(7)
Selection of the signal type to test
(1)
(7)
(6)
(5)
(2)
(3)
(4)
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......