17 Calibrating Ci conductivity sensors
17.1
Important information
17.2
General information
The calibration of Ci sensors is based on measurements in test solutions with a
defined electrolytic conductivity. Since the electrolytic conductivity of liquids is
temperature dependent, the temperature of the test solution must be sensed.
This requires either manual entry or measurement with the aid of a temperature
sensor.
17.2.1
Calibration methods for Ci conductivity sensors (inductive)
Rel. cell constant
The deviation from the nominal cell constant of a Ci sensor is described by the
relative cell constant. The rel. cell constant is determined by making a measure-
ment in a test solution with a defined conductivity.
Temperature coefficient
The temperature coefficient is a measure of the temperature dependence of the
electrolytic conductivity of a liquid. Is used to compensate for the effect of tem-
perature when measuring the electrolytic conductivity. When making a tempera-
ture-compensated conductivity measurement, the measuring value of the
conductivity is always given on the basis of the fixed, preset reference tempera-
ture. With the aid of the temperature coefficient, the value of the electrical con-
ductivity at the reference temp. is calculated from the current values of
conductivity and temperature measured for the liquid and then displayed.
Reference temperature is set in the configuration of the individual Ci analysis in-
put.
v
Chapter 10.5.4 "CR/Ci analysis inputs (conductive/inductive conductivity)",
Page 128
The temperature coefficient is determined from 2 measurements in a test solu-
tion at different temperatures (reference and operation temp.).
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
NOTE!
If the temperature coefficient of a sample solution is known, it can be entered
directly.
v
Chapter 12.2.1 "General procedure for calibration", Page 169
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......