14 Calibrating Redox sensors
186
14.3.2
Two-point Calibration
Step
Action
1
Ensure that
•
the calibration default settings are correct
•
"Percent" is set as the Redox unit in the configuration of the Re-
dox measuring input.
v
Chapter 14.2.2 "Calibration default settings for Redox sensors",
v
Chapter 10.5.3 "pH/Redox/NH3 analysis inputs", Page 126
2
Start the two-point calibration.
Device Menu
r
Calibration
r
Select Analysis Input for pH/Redox/
NH
3
r
Two-point calibration
3
Enter the concentration value of the first reference solution in per-
cent. Confirm by pressing "OK".
4
Clean the Redox electrode and immerse it in the first test solution.
Wait until the measuring value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".
5
As in step 3, enter the concentration value of the second solution
in percent. Confirm by pressing "OK".
6
Clean the Redox electrode and immerse it in the second
test solution. Wait until the measuring value displayed stabilizes
and then confirm the result of the measurement by pressing "OK".
7
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
8
Press "Yes" to accept the calibration values determined and enter
the calibration in the calibration logbook.
Press "No" to discard the results.
Tap button to enter
field
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......