background image

SUMMIT

®

 

SI  

Occipito-Cervico-

Thoracic (OCT) Spinal 

Fixation System and 

MOUNTAINEER

®

 

OCT 

Spinal System

0902-90-009  

Rev. M

Revised April 2015

© DePuy Synthes Spine, a division of DOI 2013-2015. 

All rights reserved. 

 0086

Summary of Contents for DePuy Synthes MOUNTAINEER

Page 1: ...MIT SI Occipito Cervico Thoracic OCT Spinal Fixation System and MOUNTAINEER OCT Spinal System 0902 90 009 Rev M Revised April 2015 DePuy Synthes Spine a division of DOI 2013 2015 All rights reserved 0...

Page 2: ...2 of 108 Directory en Page 3 cz Str 9 da Side 16 nl Pagina 23 fi Sivu 30 fr Page 37 de Seite 44 el 51 hu Oldal 59 it Pagina 66 no Side 73 pl Strona 80 pt P gina 87 es P gina 94 sv Sida 101...

Page 3: ...bdenum alloy conforming to ASTM F 1537 Cobalt chromium molybdenum alloy rods are intended for use with titanium components only For occipitocervicothoracic fusion either the transition rod is bent and...

Page 4: ...or with lid on to contain implant components Dry Avoid impact scratching bending or surface contact with any materials that might affect the implant surface or configuration Special attention shall be...

Page 5: ...CONCERNING TEMPORARY METALLIC INTERNAL FIXATION DEVICES WARNINGS The following are specific warnings precautions and adverse effects that should be understood by the surgeon and explained to the pati...

Page 6: ...touring of metal implants should be done only with proper equipment The operat ing surgeon should avoid any notching scratching or reverse bending of the devices when contouring Altera tions will prod...

Page 7: ...ademark of Pioneer Surgical Technology LIMITED WARRANTY AND DISCLAIMER PRODUCTS FROM DEPUY SYNTHES PRODUCTS INC ARE SOLD WITH A LIMITED WARRANTY TO THE ORIGINAL PURCHASER AGAINST DEFECTS IN WORKMANSHI...

Page 8: ...t Chromium Molybdenum P F PL FOAM Plastic Foam PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE Polyether Ether Ketone Carbon Fiber Composite PEEK OPTIMA Polyether Ether Ketone P POLYMER Polymer P CM P...

Page 9: ...Pod ln ty inky MOUNTAINEER jsou rovn dostupn v proveden ze slitiny kobalt chrom molybden vyhovuj c specifikaci ASTM F 1537 Ty inky ze slitiny kobalt chrom molybden jsou ur eny k pou it pouze s titanov...

Page 10: ...ladn ru n o istit pomoc kart e s jemn mi t tinami a istic ho prost edku s neutr ln m pH nebo znehodnotit Pokyny k i t n Pono te implant t do enzymatick ho roztoku Opl chn te jej Prove te i t n ultrazv...

Page 11: ...dchoz f z pseudoartr zou Bezpe nost a innost t chto za zen v jin ch podm nk ch nen stanovena Bezpe nostn opat en Implantaci pediklov ch roubov ch syst m pro operace p te e by m l prov d t pouze operat...

Page 12: ...a po te n m klinick m zlep en BEZPE NOSTN OPAT EN 1 CHIRURGICK IMPLANT TY NESM B T NIKDY POU ITY OPAKOVAN Explantovan implant t by nikdy nem l b t implantov n opakovan P esto e zdravotnick prost edek...

Page 13: ...oubu kter m e v st k uvoln n implant tu nebo opakovan operaci z d vodu vyjmut za zen 13 Po kozen lymfatick ch c v nebo vylu ov n lymfatick tekutiny 14 Zasa en nebo po kozen m chy 15 Fraktura struktury...

Page 14: ...l knem Si NITINOL Silikon nitinol S SS Nerezav j c ocel LOT SLO AR E REF REF KATALOGOV SLO QTY MNO STV SZ VELIKOST MADE IN VYROBENO NTI N STROJ NA NERVOVOU TK IOM N STROJE PRO NEUROMONITORING Feder ln...

Page 15: ...ov slitina titanu kobalt chrom molybdenov slitina PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE Polyetheretherketon kompozit z uhl kov ch vl ken PEEK OPTIMA Polyetheretherketon P POLYMER Polymer P C...

Page 16: ...er fremstillet af titaniumlegering Ti 6Al 4V MOUNTAINEER l ngdestave f s ogs i en legering af kobolt krom molybd n som opfylder ASTM F 1537 Stave af kobolt krom molybd nlegering er kun beregnet til an...

Page 17: ...l g p for at holde p implantatets komponenter T r Undg st d skrammer b jning eller overfladekontakt med alle materialer der kan p virke implantatets overflade eller konfiguration Der skal udvises s rl...

Page 18: ...p patienten ADVARSLER FORHOLDSREGLER OG BIVIRKNINGER VED MIDLERTIDIGE INDRE METALFIKSERINGS ANORDNINGER ADVARSLER F lgende er specifikke advarsler forholdsregler og bivirkninger som b r kendes af kir...

Page 19: ...KIRURGISKE IMPLANTATER M ALDRIG GENANVEN DES Et eksplanteret metalimplantat b r aldrig reimplanteres Selv om anordningen synes ubeskadiget kan den have sm defekter og indre stressm nstre som kan f re...

Page 20: ...tel samt mulighed for meningitis 10 Bursitis 11 Paralyse 12 L sning af skruer eventuelt f rende til l sning af implantat og eller reoperation for fjernelse af anordningen 13 Skade p lymfekar og eller...

Page 21: ...l S SS LOT PARTINUMMER REF REF KATALOGNUMMER QTY M NGDE SZ ST RRELSE MADE IN FREMSTILLET I NTI INSTRUMENT TIL NEURALT V V IOM INSTRUMENTER TIL NEUROOVERV GNING Amerikansk lov begr nser denne anordning...

Page 22: ...nium og eller kobolt krom molybd n P F PL FOAM Plast Skum PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE Polyetheretherketon Carbonfiberkomposit PEEK OPTIMA Polyetheretherketon P POLYMER Polymer P CM...

Page 23: ...le systeem bestaan uit platen moeren botschroeven staven staaf dwarsverbindingen laterale offsetverbindingsstukken koppelverbindingsstukken kabelverbindingsstukken dubbele koppelingen en axiale verbin...

Page 24: ...terilisatiemandje te laten drogen ANSI AAMI ST79 2006 Comprehensive guide to steam sterilization and sterility assurance in health care facilities Uitgebreide gids m b t stoomsterilisatie en gewaarbor...

Page 25: ...en en staven met dubbele diameter WAARSCHUWING dit hulpmiddel is niet goedgekeurd voor bevestiging van schroeven of fixatie aan de posterieure elementen pediculi van de cervicale thoracale T4 T12 of l...

Page 26: ...ook bijdragen aan vroegtijdig falen Pati nten dienen volledig te worden ge nformeerd over de risico s van falen van het implantaat 3 COMBINEREN VAN METALEN KAN CORROSIE VEROORZAKEN Er bestaan vele vo...

Page 27: ...lpmiddel 6 verhoogd risico op infectie 7 botverlies door stress shielding De chirurg dient bij de beslissing het implantaat wel of niet te verwijderen de risico s zorgvuldig tegen de voordelen af te w...

Page 28: ...INDIEN ER TUSSEN DE DATUM VAN PUBLICATIE HERZIENING VAN DEZE BIJSLUITER EN DE DATUM VAN RAADPLEGING MEER DAN TWEE JAAR IS VERSTREKEN KUNT U VOOR ACTUELE INFORMATIE CONTACT OPNEMEN MET DEPUY SYNTHES S...

Page 29: ...inium P F PL FOAM Plastic schuim PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE Polyetheretherketone koolstofvezelcomposiet PEEK OPTIMA Polyetheretherketon P POLYMER Polymeer P CM PE CoCrMo Polyethyl...

Page 30: ...dardin ASTM F 1537 mukaisesta koboltti kromi molybdeeniseoksesta valmistettuina Koboltti kromi molybdeeniseoksesta valmistetut tangot on tarkoitettu k ytett v ksi vain yhdess titaaniosien kanssa Takar...

Page 31: ...raalia puhdistusainetta tai h vitett v Puhdistusohjeet Entsyymihaude Huuhtelu Ultra nipuhdistus 10 20 minuuttia Huuhtelu Automaattinen puhdistus desinfioivassa pesukoneessa jossa on kansi implantin os...

Page 32: ...ta muiden kuin n iden tilojen hoidossa ei tunneta Varotoimi Nikamankaaren varren ruuvij rjestelm n saa implantoida ainoastaan kokemuksen omaava selk ran katoimenpiteisiin erikoistunut kirurgi joka on...

Page 33: ...upakoinnista johtuva viereisten rankaosien progressiivinen rappeutuminen voi johtaa my hemp n kliiniseen toimintah iri n toistuva kipu jopa onnistuneen fuusioitumisen ja alkuvaiheen kliinisen parantum...

Page 34: ...auksen aiheuttaa aivo selk ydinnesteen vuodon tai fistelin ja mahdollisesti meningiitin 10 Bursiitti 11 Halvaus 12 Ruuvin ulostulo mik voi johtaa implantin l ystymiseen ja tai laitteen poistamiseen ki...

Page 35: ...R KOODI REF REF LUETTELONUMERO QTY M R SZ KOKO MADE IN VALMISTUSMAA NTI NEURAALIKUDOS INSTRUMENTTI IOM NEUROMONITOROINTI INSTRUMENTIT Yhdysvaltain liitto valtion lain mukaan t m n laitteen saa myyd ai...

Page 36: ...P F PL FOAM Muovi vaahtomuovi PY Polyesteri PEEK C PEEK CARBON FIBER COMPOSITE Polyeetterieetteriketoni hiilikuitukomposiitti PEEK OPTIMA Polyeetterieetteriketoni P POLYMER Polymeeri P CM PE CoCrMo P...

Page 37: ...mini vis polyaxiales des vis monoaxiales des crochets et des c bles SONGER Les composants sont en alliage de titane Ti 6Al 4V Les tiges longitudinales MOUNTAINEER sont galement disponibles en alliage...

Page 38: ...r final pour emballer les dispositifs st ri lis s Suivre la notice d utilisation du fabricant des sachets et poches commercialis s l galement et autoris s par la FDA ou des conteneurs de st rilisation...

Page 39: ...rrecte des vis et par cons quent l utilisation de ce syst me d implants rachidiens ou de tout autre syst me Patients pouvant tre trait s de fa on s re et pr visible sans fixation interne Plaies ouvert...

Page 40: ...re souvent la fracture de fatigue des implants La quantit de compos s m talliques lib r s dans l organisme augmente galement Les dispositifs de fixation interne comme les tiges les crochets les fils...

Page 41: ...doit tre averti des limites de l implant et de la n cessit de limiter et de restreindre ses activit s physiques particuli rement les mouvements de levage et de torsion ainsi que sa participation tout...

Page 42: ...R Ba RADEL S PL SS PL Acier inoxydable plastique SRSI SS RADEL SILICONE Acier inoxydable RADEL silicone SRSN SS RADEL P F PL FOAM Plastique mousse PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE LOT...

Page 43: ...M Plastique mousse PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE Composite poly ther therc tone fibre de carbone PEEK OPTIMA Poly ther therc tone P POLYMER Polym re P CM PE CoCrMo Poly thyl ne cobal...

Page 44: ...MOUNTAINEER OCT Wirbels ulensystem bestehen aus Platten Muttern Knochen schrauben St ben Stabquerverbindungen lateralen Offset Verbindungen Kopf zu Kopf Verbindungen Kabel verbindungen Zwillingsb nder...

Page 45: ...el auch Empfehlungen f r die Trockenparameter ihrer spezifischen Ger te Nur legal vermarktete von der FDA amerikanischen Gesund heitsbeh rde zugelassene Sterilisationsfolien beutel oder wiederverwendb...

Page 46: ...weisung an den Patienten die Implantate m glichst gering zu belasten ist ein ebenso wichtiger Faktor um das Auftreten von klinischer Probleme im Zusammenhang mit dem Versagen des Fixateur Systems zu v...

Page 47: ...osionseinwirkung auf Metallimplantate ist im Allgemeinen durch das Vorhan densein von passiven Oberfl chenbeschichtungen sehr gering Ber hren sich unterschiedliche Metalle wie z B Titan und Edelstahl...

Page 48: ...l auf ein Entfernen des Implantats verzichten um die mit einer zweiten Operation verbundenen Risiken auszuschlie en 5 DER PATIENT MUSS GENAUESTENS INFORMIERT WERDEN Die postoperative Betreuung und die...

Page 49: ...SEIN FORDERN SIE BITTE BEI DEPUY SYNTHES SPINE UNTER 1 800 365 6633 ODER 1 508 880 8100 AKTUELLE INFORMATIONEN AN MATERIAL MATL LEGENDE A T ACROFLEX Ti ACROFLEX Titan A AI Aluminium A P Al PL Alumini...

Page 50: ...ium P F PL FOAM Kunststoff Schaum PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE Polyetheretherketon Kohlefaser Verbundstoff PEEK OPTIMA Polyetheretherketon P POLYMER Polymer P CM PE CoCrMo Polyethyl...

Page 51: ...CT SUMMIT SI OCT MOUNTAINEER OCT SUMMIT SI OCT MOUNTAINEER OCT SUMMIT SI OCT MOUNTAINEER SONGER Ti 6Al 4V MOUNTAINEER ASTM F 1537 OCT SUMMIT SI OCT MOUNTAINEER ISOLA MONARCH MOSS MIAMI VIPER EXPEDIUM...

Page 52: ...cal devices for reprocessing in health care facilities A guide for device manufacturers ISO 8828 AORN 132 C 4 60 ANSI AAMI ST79 2006 Comprehensive guide to steam sterilization and sterility assurance...

Page 53: ...53 of 108 OCT SUMMIT SI OCT MOUNTAINEER 3 ddd T1 T3 SONGER OCT SUMMIT SI OCT MOUNTAINEER OCT SUMMIT SI OCT MOUNTAINEER ISOLA MONARCH MOSS MIAMI VIPER EXPEDIUM 4 12 3 4 5 1...

Page 54: ...54 of 108 1 H 2 T 3 H 4 E A T B C O...

Page 55: ...55 of 108 D E E K 1 T 2 O 3 K 4 1 2 3 4 5 6 7 5 1 2 3...

Page 56: ...108 4 5 6 7 8 9 10 11 12 13 14 15 16 17 MR OCT SUMMIT SI OCT MOUNTAINEER MR OCT MOUNTAINEER MR SONGER Pioneer Surgical Technology DEPUY SYNTHES PRODUCTS INC DEPUY SYNTHES SPINE 1 800 365 6633 1 508 88...

Page 57: ...de RADEL SRTA SS RADEL Ti Al Nitride RADEL STA SS Ti Al Nitride A PY PEEK C PEEK CARBON FIBER COMPOSITE PEEK OPTIMA P POLYMER P CM PE CoCrMo P C POLYMER CARBON FIBER COMPOSITE fi Si NITINOL LOT REF RE...

Page 58: ...EL Ti Al Nitride RADEL STA SS Ti Al Nitride A S U SS ULTEM Ultem T Ti S R SS RADEL RADEL R T POLYOLEFIN RUBBER Ti T A Ti Al PY PEEK C PEEK CARBON FIBER COMPOSITE PEEK OPTIMA P POLYMER P CM PE CoCrMo P...

Page 59: ...gely csavarokb l egyten gely csavarokb l kamp kb l s SONGER huzalokb l ll Az alkatr szek tit n tv zetb l Ti 6Al 4V k sz ltek A MOUNTAINEER hosszanti rudak az ASTM F 1537 el r soknak megfelel kobalt kr...

Page 60: ...k v li haszn latuk nem enged lyezett Megtekint ssel ellen rizze hogy az implant tumon nincs e v r vagy sz vetmaradv ny Ha az implant tumon v r vagy sz vetmaradv ny l that k zileg alaposan meg kell tis...

Page 61: ...s hat konys g t csak eszk z s f zi t ig nyl jelent s mechanikai instabilit ssal s deformit ssal j r gerincbetegs gek eset ben llap tott k meg Ilyen llapotok a h ti gy ki s keresztcsonti gerincszakaszo...

Page 62: ...val kapcsolatos korl toz sokat s vint zked seket ami az implant tum el gtelenn v l s hoz vagy egy b sz v dm nyekhez vezethet C Bizonyos degenerat v betegs gek Egyes esetekben a degenerat v betegs gek...

Page 63: ...idegenanyag t l rz kenys g kialakul sa 4 Korai vagy k s i fert z s 5 A csonts r s g cs kken se az ltal hogy az implant tum tveszi a terhel st 6 F jdalom diszkomfort rz s rendellenes rzetek az eszk z...

Page 64: ...L GUSSZ M QTY MENNYIS G SZ M RET MADE IN K SZ LT NTI IDEGSZ VET ESZK Z IOM NEUROMONI TOROZ SI ESZK Z K Az USA sz vets gi t rv nyei rtelm ben az eszk z csak orvos megrendel s re rt kes thet TILOS JRA S...

Page 65: ...PY Poli szter PEEK C PEEK CARBON FIBER COMPOSITE Poli ter ter keton sz nsz las sszetett anyag PEEK OPTIMA Poli ter ter keton P POLYMER Polimer P CM PE CoCrMo Polietil n kobalt kr m molibd n P C POLYME...

Page 66: ...mponenti sono prodotti in lega di titanio Ti 6Al 4V Le aste longitudinali MOUNTAINEER sono disponibili anche in lega di cobalto cromo molibdeno ai sensi delle specifiche ASTM F 1537 Le aste in lega di...

Page 67: ...ePuy Synthes limitato ai soli Stati Uniti gli stessi non sono approvati per l utilizzo all esterno degli Stati Uniti Controllare visivamente la presenza di danni oppure di tracce di sangue o di tessut...

Page 68: ...it alcune affezioni degenerative e sensibilit ai corpi estranei Avvertenza la sicurezza e l efficacia dei sistemi spinali con viti per peduncoli sono state stabilite solo per la correzione di condizio...

Page 69: ...endere tali attivit prima della completa guarigione dell osso Anche a guarigione completata possibile che il paziente non sia in grado di riprendere completamente tali attivit B Senilit disturbi menta...

Page 70: ...RALI 1 Disunione o unione tardiva 2 Piegamento o rottura dell impianto Allentamento dell impianto 3 Sensibilit ai metalli o reazione allergica a un corpo estraneo 4 Infezione precoce o tardiva 5 Dimin...

Page 71: ...FIBER COMPOSITE LOT NUMERO LOTTO REF REF NUMERO DI CATALOGO QTY QUANTIT SZ DIMENSIONI MADE IN PRODOTTO IN NTI STRUMENTO PER TESSUTI NEURALI IOM STRUMENTI PER NEUROMONITORAGGIO La legislazione federale...

Page 72: ...P F PL FOAM Plastica gomma spugnosa PY Poliestere PEEK C PEEK CARBON FIBER COMPOSITE Polietereterchetone Composto di fibra di carbonio PEEK OPTIMA Polietereterchetone P POLYMER Polimero P CM PE CoCrM...

Page 73: ...1537 Legeringsstengene av kobolt krom molybden skal kun brukes sammen med komponenter av titan For occipitocervicotorakalfusjon skal overf ringspinnen enten b yes og kappes til passende lengde eller...

Page 74: ...atisk rengj ring i et desinfiseringsapparat med lokk som holder implantatkomponentene p plass T rking Unng st t skrammer b ying eller overflatekontakt med materialer som kan p virke implantatoverflate...

Page 75: ...de p pasienten ADVARSLER FORHOLDSREGLER OG UHELDIGE VIRKNINGER VEDR RENDE MIDLERTIDIGE INDRE FIKSASJONSANORDNINGER AV METALL ADVARSLER Det f lgende er spesielle advarsler forholdsregler og motreaksjon...

Page 76: ...lsen FORSIKTIGHETSREGLER 1 KIRURGISKE IMPLANTATER M ALDRI BRUKES P NYTT Et eksplantert implantat skal aldri implanteres p nytt Selv om en anordning ser ut til v re uskadet kan den ha sm defekter og in...

Page 77: ...reparere skader kronisk cerebrospinalv ske lekkasje eller mulig meningitt 10 Bursitt 11 Paralyse 12 Skruer l sner noe som kan f re til at implantatet l sner eller til ny operasjon for fjerne anordnin...

Page 78: ...ALOGNUMMER QTY ANTALL SZ ST RRELSE MADE IN PRODUSERT I NTI NERVEVEVS INSTRUMENT IOM INSTRUMENTER TIL NEVROLOGISK OVERV KING I henhold til amerikansk lov kan dette produktet kun selges av lege eller p...

Page 79: ...P F PL FOAM Plast skum PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE Polyetereterketon karbonfiberkompositt PEEK OPTIMA Polyetereterketon P POLYMER Polymer P CM PE CoCrMo Polyetylen kobolt krom mol...

Page 80: ...e pr ty MOUNTAINEER wykonane ze stopu kobaltowo chromowo molibdenowego spe niaj cego norm ASTM F 1537 Pr ty ze stopu kobaltowo chromowo molibdenowego przeznaczone s wy cznie do u ytku z elementami tyt...

Page 81: ...zacji wielokrotnego u ytku firmy DePuy Synthes mo na stosowa wy cznie w Stanach Zjednoczonych i nie zosta y zatwierdzone do u ycia poza terytorium Stan w Zjednoczonych Sprawdzi wzrokowo pod k tem uszk...

Page 82: ...liwo zrostu tj rak dializa lub osteopenia oty o pewne choroby zwyrodnieniowe i wra liwo na cia a obce Ostrze enie Bezpiecze stwo i efektywno zestaw w do kr g w opartych na rubach szypu kowych zosta o...

Page 83: ...ania Te stany w r d wielu innych mog sprawi i pacjent b dzie ignorowa pewne wa ne ograniczenia i rodki ostro no ci zwi zane z u yciem implant w prowadz c do uszkodzenia lub innych komplikacji C Pewne...

Page 84: ...ne nadwyr eniem 6 B l dyskomfort lub nietypowe odczucia zwi zane z obecno ci przyrz du 7 Uszkodzenie nerwu zwi zane z urazem operacyjnym lub obecno ci przyrz du Trudno ci neurologiczne w tym b l korze...

Page 85: ...UKOWANO W NTI NARZ DZIE DO TKA NEK NERWOWYCH IOM INSTRUMENTY DO NEUROMONITORINGU Zgodnie z prawem federalnym USA urz dzenie to mo e by sprzedawane wy cznie przez lub na zlecenie lekarza NIE PODDAWA PO...

Page 86: ...ibden P F PL FOAM Tworzywo sztuczne Pianka PY Poliester PEEK C PEEK CARBON FIBER COMPOSITE Polieteroeteroketon W kno w glowe PEEK OPTIMA Polieteroeteroketon P POLYMER Polimer P CM PE CoCrMo Polietylen...

Page 87: ...e fixa o miniparafusos poliaxiais parafusos monoaxiais ganchos e cabos SONGER Os componentes s o fabricados em Liga de Tit nio Ti 6Al 4V As hastes longitudinais MOUNTAINEER tamb m est o dispon veis em...

Page 88: ...os terminalmente As instru es de utiliza o do fabricante para o inv lucro de esteriliza o bolsas ou recipientes de esteriliza o reutiliz veis da DePuy Synthes devem ser seguidas A utiliza o dos recipi...

Page 89: ...clui se qualquer entidade ou doen a que impe a totalmente a possibilidade de fus o por exemplo neoplasias malignas di lise renal ou osteopenia obesidade algumas doen as degenerativas e sensibilidade a...

Page 90: ...ceberem dispositivos de fixa o interna os factores que se seguem podem ser extremamente importantes para o eventual sucesso do procedimento A Profiss o ou actividade do doente Caso a profiss o ou acti...

Page 91: ...nte dever compreender que um implante met lico n o t o forte como um osso saud vel normal e que se poder soltar dobrar e ou partir se for sujeito a um esfor o excessivo especialmente na aus ncia de um...

Page 92: ...COMPOSITE Poli ter ter cetona Composto de Fibra de Carbono PEEK OPTIMA Poli ter ter cetona LOT N MERO DE LOTE REF REF N MERO DE CAT LOGO QTY QUANTIDADE SZ TAMANHO MADE IN FABRICADO EM NTI INSTRUMENTO...

Page 93: ...l stico Espuma PY Poli ster PEEK C PEEK CARBON FIBER COMPOSITE Poli ter ter cetona Composto de Fibra de Carbono PEEK OPTIMA Poli ter ter cetona P POLYMER Pol mero P CM PE CoCrMo Polietileno Cobalto Cr...

Page 94: ...s y cables SONGER Los componentes est n fabricados con aleaci n de titanio Ti 6Al 4V Las varas longitudinales MOUNTAINEER tambi n est n disponibles en aleaci n de cobalto cromo molibdeno conforme a la...

Page 95: ...ontenedores de esterilizaci n reutilizables de DePuy Synthes solo se limita a los Estados Unidos y no se aprueban para su uso fuera de ese pa s Inspeccione visualmente el implante para detectar da os...

Page 96: ...renal u osteo penia obesidad ciertas enfermedades degenerativas y sensibilidad a cuerpos extra os Advertencia la seguridad y eficacia de los sistemas espinales de tornillo pedicular s lo se han compr...

Page 97: ...caminar mucho o realizar tareas manuales no deber retomar sus actividades hasta que el hueso haya cicatrizado completamente Incluso despu s es posible que el paciente no pueda retomar sus actividades...

Page 98: ...ue se salen de su lugar o sufren da os por la realizaci n de actividades inadecuadas pueden desplazarse y da ar los nervios o los vasos sangu neos Los pacientes activos debilitados o con las facultade...

Page 99: ...r cetona fibra de carbono PEEK OPTIMA Poli ter ter cetona P POLYMER Pol mero P CM PE CoCrMo Polietileno LOT N MERO DE LOTE REF REF N MERO DE CAT LOGO QTY CANTIDAD SZ TAMA O MADE IN FABRICADO EN NTI IN...

Page 100: ...Cromo Molibdeno PY Poli ster PEEK C PEEK CARBON FIBER COMPOSITE Compuesto de poli ter ter cetona fibra de carbono PEEK OPTIMA Poli ter ter cetona P POLYMER Pol mero P CM PE CoCrMo Polietileno Cobalto...

Page 101: ...ngar kabelf sten dubbelringar axialkopplingar inst llningsskruvar minipolyaxiala skruvar monoaxiala skruvar krokar samt SONGER kablar Komponenterna r tillverkade av en titanlegering Ti 6Al 4V MOUNTAIN...

Page 102: ...och ett pH neutralt reng ringsmedel eller kasseras Reng ringsinstruktioner Enzymatisk bl tl ggning Sk ljning Ultraljudsreng ring 10 20 minuter Sk ljning Automatiserad reng ring i en tv ttdesinfektor m...

Page 103: ...ar f r andra tillst nd r inte k nda F rsiktighets tg rd Implantation av spinala pedikelskruv system skall endast utf ras av erfarna ryggkirurger som genomg tt speciell utbildning i anv ndningen av det...

Page 104: ...fus degeneration i inter vertebraldiskar Progressiv degeneration i intilliggande segment orsakad av r kning kan leda till sen klinisk svikt recidiv av sm rta ven efter lyckad fusion och klinisk f rb t...

Page 105: ...en 9 Durarifter som uppst r under operationen kan resultera i behov av ytterligare kirurgi f r reparation av duran kroniskt l ckage av cerebrospinalv tska eller fistel samt meningit 10 Bursit 11 F rla...

Page 106: ...PARTINUMMER REF REF KATALOGNUMMER QTY ANTAL SZ STORLEK MADE IN TILLVERKAD I NTI NERVV VNADS INSTRUMENT IOM INSTRUMENT F R NEUROMONITORERING Enligt amerikansk federal lagstiftning f r denna anordning e...

Page 107: ...olybden P F PL FOAM Plast skumplast PY Polyester PEEK C PEEK CARBON FIBER COMPOSITE Polyetheretherketone kolfiberkomposit PEEK OPTIMA Polyeter eterketon P POLYMER Polymer P CM PE CoCrMo Polyetylen kob...

Page 108: ...ersteller Az elismert gy rt val kapcsolatban l sd a term kc mk t Per conoscere il fabbricante legale fare riferimento all etichetta del prodotto Aktuell produsent er angitt p produktetiketten Informac...

Reviews: