RTC - Real Time Check
21
General
This function allows you to program up to three different times for the device to
carry out a self-diagnosis of defibrillation, battery level, connected paddles, and
verification of the device's connection to the power outlet.
RTC Operation on DualMax
The configurations for the RTC function can be accessed through the menu
CONFIGURATION > RTC.
The configurations screen of the RTC function allows you to define the number of
automatic daily tests to be performed, configurable in values from 1 to 3 tests. It is
also possible to configure the time for each of these daily tests, carry out a manual
test, or check the status of the last test carried out.
Imprimir relatorio ultimo teste
1
2
3
8
9
5
4
7
6
1 - Number of daily tests
Sets the amount of daily automatic tests. It can be configured with values from 1 to
3 tests per day.
2 - Test 1 time
Schedule of the first daily test. Configured by the user.
165