User manual |
Monitor mode - ECG
106
3 - Detect pacemaker
Enables to turn the pacemaker detection mode on or off. When on, the equipment
indicates on the screen the pacemaker’s pulse moment. Detect pacemaker mode
must only be used on patients with pacemaker.
4 - Amplitude
Selects the gain of the ECG amplification stage. Selectable in 5, 10, 15, 20, 30, 40
mm/mV.
5 - Frequency bandwidth selection
Filter selection for power interference. “Diag” or “Monitor”.
6 - Mains supply filter
Enables to turn filter on or off to reduce mains supply interference with the ECG
signal.
7 - ST
Turns on or off the ST segment monitoring.
8 - Sudden Death Prevention (SDP)
Allows to turn on or off the Sudden Death Prevention mode. When turned on,
in monitoring through the adhesive pads or electrodes, the equipment alarms
and indicates “shock indicated” if a ventricular fibrillation/tachycardia (VF/VT) is
identified.
9 - Next page
Displays the next screen of the ECG menu.
10 - Back/Exit
“Back” to Configuration menu and “Exit” to monitoring screen.