Alarms and limits
62
CO
2
: it is possible to adjust the minimum CO
2
alarm to levels between 18 and
96 mmHg, with intervals of 3 mmHg in "adult" mode and intervals of 1 mmHg
in "neo" mode. It is possible to set the maximum CO
2
alarm to levels between
21 and 99 mmHg, with intervals of 3 mmHg in "adult" mode and intervals of 1
mmHg in "neo" mode.
RESP: it is possible to adjust the minimum respiration alarm to levels
between 3 and 147 rpm, with intervals of 3 rpm in "adult" mode and intervals
of 1 rpm in "neo" mode. It is possible to set the maximum respiration alarm
to levels between 6 and 150 rpm, with intervals of 3 rpm in "adult" mode and
intervals of 1 rpm in "neo" mode.
Technical alarm test
To perform the alarm test do the following steps:
1 - Turn the equipment on, without cables or sensors connected. It should occur
indication of technical alarm (text messages in Infocenter) disconnected sensor.
2 - After confirming the visual indication, connect the sensor correctly. If the
indication disappears, the alarm is working properly.
Otherwise, change the sensor and repeat the test. If the result is the same, the
alarm will probably be defective.
3 - Repeat this procedure for the other modules, remembering to test with their
respective sensor and parameters.
4 - With the indication of an active technical alarm, press the button PAUSE AUDIO
for 1s and check on screen the suspended alarm indication for all parameters.
Wait for 60 s and the alarm will activate. On the screen, the suspended alarm signal
disappears, indicating the return of the alarm sound.
The time the alarm stays turned off is 60 s.
To turn the alarm sound on again, press the PAUSE AUDIO button for 1 second.
The sound of the parameters of alarm can be turned on and off individually in the
"Alarm" menu and in parameter menus.