User manual |
Specifications and safety
214
Tachycardia alarm time:
206 BPM (1 mV): 5 seconds.
206 BPM (0.5 mV): 5 seconds.
206 BPM (2 mV): 5 seconds.
195 BPM (2 mV): 5 seconds.
195 BPM (1 mV): 5 seconds.
195 BPM (4 mV): 5 seconds.
NIBP - Non-Invasive Arterial Pressure (optional)
Technique:
Oscillometric.
Armband:
Use only armband that has connector according to norm
ISO 594-1.
Automatic mode:
1, 2, 3, 4, 5, 10, 15, 30, 60 e 90 minutes.
Manual mode:
One measurement.
Adult range:
Systolic: 40 to 260 mmHg.
Mean: 26 to 220 mmHg.
Diastolic: 20 to 200 mmHg.
Pediatric range:
Systolic: 40 - 160 mmHg.
Mean: 26 - 133 mmHg.
Diastolic: 20 - 120 mmHg.
Neonatal range:
Systolic: 40 to 130 mmHg.
Mean: 26 to 110 mmHg.
Diastolic: 20 to 100 mmHg.
Overpressure limit per
software:
Adult: 290 mmHg max.
Neonatal: 145ƒ mmHg max.
Overpressure protection by
hardware:
Adult: 300 ± mmHg.
Neonatal: 150 ± mmHg.
Resolution:
1 mmHg.