Alarms and limits
51
• Visual frequency: 2 Hz.
• Frequency of verification of the physiological alarms system by the user:
monthly.
STANDARD ALARM VALUE
The following tables provide the factory default values for the alarm limits:
ECG
NIBP
Bpm Max.
160 BPM
Medium Min.
50 mmHg
Bpm Min.
40 BPM
Initial pressure
140 mmHg
Capnography
Respiration
CO
2
Max.
60 mmHg
Max.
45 RPM
CO
2
Min.
21 mmHg
Min.
6 RPM
SpO
2
ST segment
Answer.
Normal
Max.
10 mm
SPO
2
Max.
100%
Min.
-10 mm
SPO
2
Min.
85%
Pulse Max
.
160
IP (channels 1 and 2)
Pulse Min.
40
Systole Max.
200 mmHg
Systole Min.
90 mmHg
Temperature (channels 1 and 2)
Diastole Max.
150 mmHg
Max.
38.0°C
Diastole Min.
50 mmHg
Min.
30.0°C
Average Max.
170 mmHg
Average Min
.
50 mmHg
Technical alarm (medium priority)
Sound and visual signals indicate that the DualMax is not able to accurately monitor
the patient's status. The technical alarm indications are shown in the information bar
(see item "Information bar").
In addition to the technical alarm conditions indicated in the information bar, the
are two more conditions: "bad contact" and "battery charge level" (see following
sections).