User manual |
PC connection
175
RTC software installation
• Insert the program CD in the CD/DVD ROM drive.
• If the installer does not initiate automatically, locate the "RTC" file or "Real Time
Check" on the program CD and double click on it.
• Follow the installation instructions that appear on the screen.
NOTE: check that the "Wireless Reception Unit" is connected to the
computer by USB prior to installing the software. After completing the
installation, links will be created to run the program on the Windows start
menu and on the desktop.
RTC software operation
The RTC software runs in the background, with the minimized system awaiting a
device to communicate with the "Real Time Check Control Center".
The DualMax can be configured to perform self tests at times specifically configured
by the user. The device will send the results of the following tests to the RTC
system:
• Test of the connection with the ECG board.
• Test of the defibrillator module.
• Check of the battery level.
• Check of the connection with the power outlet.
Upon receiving the test results, the RTC software will open a screen showing
this data. This information will also be stored in a log storage file. Optionally, the
program can be set up to send e-mails with this data to determined recipients.
After the program is run for the first time, the software will be configured to
automatically run after Windows starts up. The RTC icon is in the shape of a heart
and will be shown on the task bar, next to the clock of the operating system, as
shown in the figure below: