User manual |
Specifications and safety
212
Electrode error:
NO ELECTRODES and a trace line will appear on the
display if there is a disconnected cable or electrode.
Low ECG amplitude or
saturated ECG signal:
The “Searching for ECG Signal” message will appear on the
screen or printer (electrocardiograph mode) when the ECG
10mm/mV amplitude is less than 2.4 mm peak-to peak (or
sensitivity proportional).
Multifunction pads error:
If an adhesive pad is disconnected, a trace line will appear
on the display.
Range:
Precision:
15 to 350 BPM.
± 1 BPM of 15 to 350 BPM.
Rejection in common mode:
Higher than 90 dB, measurement in compliance with the
AAMI standards for heart monitors (EC 13).
Sensitivity:
5, 10, 15, 20, 30 and 40 mm/mV.
Scan speed:
12.5, 25 or 50 mm/s.
Filter:
35 Hz or 60 Hz.
ECG response frequency:
Diagnostic mode (0.05 - 100 Hz).
Monitor mode (1 - 40 Hz).
Patient isolation’s
(defibrillation proof):
ECG: type CF.
SpO
2
: type CF.
Physiological alarm:
Alarm not locked.
Minimum level (30-100).
Maximum level (100-250).
Visual indication.
Sound indication.
Technical alarm:
Alarm not locked.
Visual indication.
Sound indication.