Alarms and limits
55
SpO
2
- Disconnected
sensor.
MEDIUM PRIORITY
Yellow
Sensor or SpO
2
extension disconnected or
incorrectly placed.
SpO
2
- Artifact.
MEDIUM PRIORITY
Yellow
Muscle tremor detected.
Weak signal.
MEDIUM PRIORITY
Yellow
DualMax cannot identify the signal. Weak signal,
possibly patient with low perfusion.
Loss of pulse.
HIGH PRIORITY
RED
Patient without heartbeats for more than 4
seconds.
(↑) SPO
2
- Sat
above the limit.
HIGH PRIORITY
RED
Saturation is above the value defined on the
alarm configuration.
(↓) SPO
2
- Sat
below the limit.
HIGH PRIORITY
RED
Saturation is below the value defined on the
alarm configuration.
(↑) SPO
2
- BPM
above the limit.
HIGH PRIORITY
RED
Pulse value is above the value defined in the
alarm configuration.
(↓) SPO
2
- BPM
below the limit.
HIGH PRIORITY
RED
Pulse value is below the value defined in the
alarm configuration.
NIBP
Alarm
Priority
Color
Possible cause
Excessive
pressure.
MEDIUM PRIORITY
YELLOW
The cuff´s maximum pressure has been
exceeded.
Cuff problems.
MEDIUM PRIORTY
YELLOW
The cuff is misplaced or there is leakage on
measure circuit.
Weak signal .
MEDIUM PRIORITY
YELLOW
Pulse too weak for NIBP measurements. Check
the cuff´s position and tightening.
Excessive
movement.
MEDIUM PRIORITY
YELLOW
There is noise due to the patient’s movement.
Long
measurement.
MEDIUM PRIORITY
YELLOW
The measurement of pressure is too long with
possibility of inaccurateness.
(↑) NIBP - Sys
above the limit.
HIGH PRIORITY
RED
Systolic pressure value is above the value set in
the alarm setting.
(↓) NIBP - Sys
below the limit.
HIGH PRIORITY
RED
Systolic pressure value is below the value set in
the alarm setting.
(↑) NIBP - Dia
above the limit.
HIGH PRIORITY
RED
Diastolic pressure value is above the value set
in the alarm setting.
(↓) NIBP - Dia
below the limit.
HIGH PRIORITY
RED
Diastolic pressure value is below the value set in
the alarm setting.