The essential performance of the DualMax is understood as the ANALYSIS of the
patient's heart rate, the DEFIBRILLATION, CARDIOVERSION and PACEMAKER
applied to the patient, and the monitoring of the PATIENT in electrocardiogram,
oximetry, non-invasive pressure, respiration, and invasive pressure, The
performance of the DualMax is designed and verified for the absence of an
unacceptable risk.
If performance is lost or degraded due to electromagnetic disturbances:
• The ECG signal may be interfered with and cardiac rhythm analysis may be
compromised as long as electromagnetic disturbances persist.
• The SPO2 curve can be disturbed, besides the possibility of alteration in the
measured value.
• Non-invasive pressure measurement may suffer alterations on the measured
values.
• The Invasive pressure curve may suffer interferences.