User manual |
Defibrillator mode
74
During the equipment's charging cycle, the value of the energy that has already
been stored is displayed in RED. When the charge is complete the numbers are
displayed in RED and blink, indicating the equipment is ready and the shock can be
delivered.
Charge Auto-Sequencing mode (Auto Seq)
In the Auto Seq mode, the shock energy
levels will follow the order previously
set by the user in: "Defibrillation >
AUTO SEQUENCING CHARGE” (See
“Defibrillation setup” below).
There are three sequential energy levels,
being that, from the third shock on all
subsequent shocks will use the same
discharge value of the latter. The preset
sequence will be interrupted in the following
conditions:
• Equipment shutdown.
• Auto Seq mode exit.
• Reconfiguration of the Auto-Sequencing mode energy levels (available after the
delivery of the third shock).
In case child pads are used, the equipment will automatically limit the charge value
to 50 J.
The energy to
be charged is
indicated in red.
After shock
delivery, the
energy is
updated with
the next energy
level.