Removable battery
Even when disconnected (stand-by), the DualMax executes internal routines
checking the status of the equipment. In spite of this procedure entailing a low
power consumption, the battery charge may be consumed. Therefore, whenever
the device has not been connected to an electric current for more than 20 days, it is
advisable to execute a full battery charge. If this procedure is not performed, there
is a risk of draining the battery and consequently being unable to use the DualMax
in its portable configuration (not connected to the electric current). To recharge the
battery, connect the monitor to charge the battery, connect the monitor to an AC
power source (110 or 220 V outlet) or a DC power source.
There are no restrictions or limitations for using the DualMax while its battery is
being recharged by an AC source or DC External source.
Every battery has a determined lifetime, which is the possible quantity of full charge
and discharge cycles, without loss of performance (see battery specifications in
chapter 8). If the equipment presents a loss in battery performance, please request
a new set to Instramed's technical assistance, To request pieces and services,
please contact Instramed at +55 (51) 3073 8200.
Battery replacement is recommended every 02 (two) years or when the runtime is less
than 01 (one) hour.