Defibrillator mode
72
Step 3 - Shock
After the “Charge Ready” warning, press the "Shock" 3 button (orange) in the front
panel or use
the two buttons
(orange) in the external paddles.
It is only possible to defibrillate using the pad buttons with the adult/child external
pads.
CAUTION: make sure nobody is touching the patient! Tell passersby to stand
clear of the patient!
The number of shocks and length of operation are indicated on the screen.
Synchronism - Synchronized discharge - Cardioversion
Remember: the function “Synchronized Shock”
is disabled after the shock is delivered.
Monitor the patient with 3 or 5-leads ECG cables or with the defibrillation electrodes.
Press the “Sync” button in the front panel. Ensure that the synchronization marker
is red and lined up with the 'R” wave and the “SYNC” indication is displayed next to
the selected energy value.
Follow steps 1-2-3 for defibrillation.
IMPORTANT: Keep key 3 (shock) or the two shock pads’ buttons pressed until
the next “R” wave is identified. The DualMax will deliver the shock when the
next “R” wave is identified.
IMPORTANT: If DualMax does not identify a valid QRS it will not trigger the
shock! For this reason, do not use the cardioversion in Ventricular Fibrillation
rhythms.
Cardioversion can cause discomfort to the patient, as well as skin redness.