User manual |
Specifications and safety
211
Frequency:
Between 30 ppm to 180 ppm (increments of 5 ppm),
precision ± 2%.
Refractory period:
340 ms (between 30 to 80 ppm).
240 ms (between 90 to 180 ppm).
Maximum output voltage:
350 V.
ECG
Input:
ECG patient cable and the corresponding derivations:
- 03 leads (optional): DI, DII and DIII.
- 05 leads: DI, DII, DIII, aVL, aVR, aVF and C.
- 10 leads (optional): DI, DII, DIII, aVL, aVR, aVF, C1, C2,
C3, C4, C5, C6.
External defibrillation pads.
Multifunctional adhesive pads.
Supports up to 12 simultaneous derivations when equipped
with the ECG optinal of 12 derivations.
ECG electrodes (adult or
child):
COMPOSITION: adhesive conductor with hydrogel and
silver sensor (Ag/AgCl), latex free.
USE INSTRUCTIONS: 1) Shave hairs, clean the spot and
dry with gauze. 2) Before removing the protective film,
attach the ECG wire to the electrode. 3) Remove the film
from the electrode and put it in place by pressing firmly.
PRECAUTIONS: do not open the packaging before
use. Do not apply the electrodes on the nipples, bony
protuberance, cuts, lesions, scars, only on intact skin. Store
at temperatures between 15°C and 40°C. Non-sterile
single-use product. Discard after use.
ECG gel:
It is recommended the use of a gel suitable for medical
procedure, indicated for use in electrocardiography,
defibrillators and similar. Used as a contact medium for
the transmission of electrical impulses by means of an
electroconductor. Biocompatible. After each use, clean the
accessories according to cleaning recommendations.