User manual |
Monitor mode - NIBP
111
Monitoring Non-Invasive Pressure
1 -
Connect the end of the extension hose to the equipment’s non-invasive pressure
connector.
2 -
Measure the limb in which the armband will be applied to and select an
adequate type. See the chart below.
3 -
Place the armband on the limb as it suggests “
Armband positioning”.
4 -
Connect the armband to the extension hose.
5 -
Select one of the measurement modes: manual or automatic.
ARMBAND SELECTION
Armband
Limb circumference (arm/leg)
Child
10 to 19 cm
Pediatric
18 to 26 cm
Adult
25 to 35 cm
Extra Large
33 to 47 cm
PLACING THE ARMBAND
1 - Select the measurement place. Choose a place with good blood circulation,
without skin problems and in which armband use does not harm the patient.
Because of convenience and because of the fact that standard values are based on
this location, give preference to the upper arm.
2 - Check the armband’s appropriate size for the chosen area according to the
previous chart.
3 - Make sure the limb is supported to
guarantee that the armband is at the
heart’s level. Due to the hydrostatic effect,
placement above or below the heart’s level
may cause incorrect measurements.
4 - Ensure the ARTERY Mark is above the
brachial artery.