User manual |
Alarms and limits
61
After editing an alarm, to return to the alarms menu, only select the button
present on the alarm configuration screens of each parameter.
Minimum / Maximum limits adjustment
The adjustment of minimum and maximum values is done individually on each
parameter by using the e-Jog Control. The operator must first select the limit and
the parameter to be modified and then press it. Next, the desired value must be
adjusted and then pressed again.
ECG: it is possible to adjust the minimum ECG alarm to levels between 30 and
100 BPMs with intervals of 5 BPMs in “adult” mode and intervals of 1 BPMs
in “neo” mode. It is possible to adjust the maximum ECG alarm to levels
between 100 and 250 BPMs with intervals of 5 BPMs in “adult” mode and
intervals of 1 BPMs in “neo” mode.
SpO
2
: it is possible to adjust the minimum SpO
2
alarm to levels between
40 and 99 BPMs with intervals of 5 BPMs in “adult” mode and intervals of
1 BPMs in “neo” mode. It is possible to adjust the maximum ECG alarm to
levels between 41 and 100 BPMs with intervals of 5BPMs in “adult” mode and
intervals of 1 BPMs in “neo” mode.
NIBP: it is possible to adjust the minimum NIBP alarm to levels between 50
and 290 mmHg to systolic, diastolic and mean pressure with intervals of 5
mmHg. It is possible to adjust the maximum NIBP alarm to levels between 60
and 300 mmHg to systolic, diastolic and mean pressure with intervals of 5
mmHg.