User manual |
RTC - Real Time Check
166
3 - Test 2 time
Schedule of the second daily test. Configured by the user. This item is disabled
when the number of daily tests is 1.
4 - Test 3 time
Schedule of the third daily test. Configured by the user. This item will be disabled
when the number of daily tests is less than 3.
5 - Perform manual test
Performs a manual test, with no prior configuration required.
6 - Last test
Displays data from the last test performed or the current step when a test is running.
7 - Show last test report
Shows the results of the last test performed.
8 - Print last test report
Prints the results of the last test performed.
9 - Back/Exit
"BACK" to the configuration menu or "EXIT" to the monitoring screen.
Automatic test
To perform the automatic test, the equipment should be turned off at the
programmed time.
At the configured time, the DualMax will automatically turn on and begin the test,
composed of the following steps:
• Test of the connection with the ECG board.
• Test of the defibrillator module.
• Check of the battery level.
• Check of the connection with the power outlet.