User manual |
Specifications and safety
204
Synchronized command:
Sync key in front panel.
Charge indicators:
Audio indication of equipment being charged.
Audio indication of charge completed.
LED on external pads and charge level indicated on display.
Maximum charging time:
200 J: mains and battery < 4 s.
360 J: mains and battery < 6 s.
Pads (options):
Adult and child external (included).
Multifunctional for pacemaker, monitoring and defibrillation
(optional).
Multifunction extension (optional).
NOTE: the adhesive pads provided by Instramed have a
biocompatibility certificate in accordance with ISO 10993.
External pads size:
Adult = 10.3 cm x 8.5 cm (contact area = 81.9 cm²).
Children = 4.5 cm x 4 cm (contact area = 18 cm²).
Pads cable length:
2 meters.
Cardioversion:
< 60 ms.
Maximum output voltage:
2000 V.
Maximum output electric
current:
70 A (25 Ω).
AED mode (optional)
Functional characteristics:
Voice commands, visual indications, CPR instructions,
USB 2.0 for PC connection, multi-languages, SDP (Sudden
Death Prevention) technology.
USB:
USB 2.0 for tranfering electrocardiogram stored in AED
mode to a compatible PC.
SoftDEA:
Software for viewing data tranferred to PC.