User manual |
Pacemaker mode
99
Starting stimulation
1 -
If it not yet connected, connect the adhesive pads cable to the DualMax.
2 -
Check if the multifunction adhesive pads package is intact and within the
expiration date.
3 -
Insert the adhesive pads connector into the equipment's extension cable.
4 -
Put the adhesive pads on the patient according to package instructions.
5 -
In case of demand mode stimulation, apply ECG monitoring electrodes.
Fixed stimulation
1 -
Rotate the dial to the pacemaker mode.
The pacemaker starts in PAUSE, without stimulation pulses.
The mode must be changed to FIXED.
2 -
Select the leads for ECG viewing.
3 -
With the e-Jog, adjust the initial current and frequency (PPM) values (see the
previous page). The current value must be as low as possible.
4 -
Go to “ON” with the e-Jog, in order to start stimulation. A message in the
“Infocenter” will inform the pacemaker is active.
5 -
Check if the pacemaker's pulse meter appears on the screen.
6 -
Increase the stimulation current until the heart rate is captured. This is indicated
by the presence of the QRS complex (Q, R and S waveforms) right after the
pacemaker marker.