Minifors 2 - Benchtop Bioreactor
Operation
07 February 2019
Page
139
of
214
1) In the parameter group that contains the parameter in ques-
tion, a number on a red background appears. This indicates
the number of existing parameter alarms.
2) The parameter in question is displayed with a red bar and a
red actual value.
3) A red exclamation mark highlighted in white on a red back-
ground appears in the footer.
Pressing the symbol or swiping upwards opens the
Equipment
Alarm
menu. For details, see the “Alarms – Equipment Alarm
Menu” section.
Parameter alarms are also logged in the batch log file, see the
“USB data export and import from a USB stick” section.
Parameter alarm limit factory settings
Parameters
Alarm limits
Temperature
2 °C
Stirrer
50 min
-1
pH
0.5
pO2
10 %
TotalFlow
0.3 L min
-1
GasMix
10 %
AirFlow
0.3 L min
-1
Gas2Flow
0.3 L min
-1