Minifors 2 - Benchtop Bioreactor
Before Cultivation
07 February 2019
Page
97
of
214
8.1.24 Setting the Pumps
If the pumps are not used with the default settings, we recommend
that the appropriate settings are now made on the operating panel.
It is possible, for example, to estimate and display the volume (in
mL) that has been pumped since the Batch (process) started. To
this end, the diameter of the hose used must be selected.
For details on the pumps and the setting options, see the main
chapter “Operation”, the “PUMPS parameter group” chapter and
the associated sub-chapters.
8.1.25 Removing the Pump Heads
To remove the pump heads from the basic unit, proceed as fol-
lows:
Procedure
1.
Swing open the pump cover.
2.
Remove the mounting plate with the pump heads from the
drive shafts by holding the two handles.
3.
Place the mounting plate with the pump heads onto the pump
holder on the vessel holder.