Screw the sensor on the hollow screw into the port and tighten
it.
10.
Tighten the grub screw in the support guide with the key.
8.1.19 Mounting the Antifoam Sensor
Please note the following points for mounting:
The antifoam sensor is equipped with transparent insulation
that must be intact, as otherwise a continuous signal
“Foam/liquid detected” may be generated.
The sensor head must not touch the clamping adaptor, other-
wise a continuous short-circuit is generated, indicating
“Foam/liquid detected”.
The clamping adapter on the sensor must be equipped with
an intact O-ring.
Proceed as follows for mounting:
Procedure
1.
Remove the protective cap from the sensor.
2.
Insert the sensor into the port.