Minifors 2 - Benchtop Bioreactor
Accessories
07 February 2019
Page
57
of
214
sample has to be processed under sterile conditions. For sampling,
the same non-sterile syringe can be used repeatedly, without fear
of contamination of the culture vessel.
Aseptic Sampling
For each sample, use a new, sterile syringe with Luer Lock fitting,
in order to ensure the sterility of the sample.
Sterile syringes are consumables and therefore not included in the
set.
INFORMATION
The use of another syringe is also possible. But a syringe with
Luer lock prevents unwanted movement of the syringe.
Before fitting the syringe, disinfect the sample valve. Fort this,
spray a commercially available disinfectant onto the valve.
After spraying and after each sampling, close the the sample
valve with a sterile Luer-Lock cap (Dead End Cap) to keep the
valve and sample sterile.
The caps are not included in the kit. Very convenient to use are so-
called combi-caps that fit on male and female connectors alike.
Caps that are vented and made of steam sterilisable material can
also be fitted during autoclaving.
5.14 Pump Heads
The autoclavable pump heads are fitted with PharMed pump hoses
prior to delivery. Three different hose diameters are available for
different delivery rates:
1.0 mm (standard)
0.5 mm
2.5 mm
For more detailed information about pumps and hoses refer to
main chapter “Technical Data”, chapter “Specification”, “Pumps