Minifors 2 - Benchtop Bioreactor
Operation
07 February 2019
Page
175
of
214
If the calibration was successful, the menu display changes
automatically to calibrate the second point.
10.
In the pull-down-menu, select the value 100 (%).
The selection is displayed.
CALIBRATE POINT 2 (or 1, if not calibrated one after the
other in 1-point calibration mode)
is activated, i.e. the but-
ton turns orange.
INFORMATION
This facilitates a calibration of the sensor outside of the medium,
i.e. without active gassing for a standard calibration outside the
medium, which is not described here.
11.
Enter the setpoint 100 (%) in the input field.
The input field turns orange and displays the set setpoint.
Gassing with nitrogen is activated (if connected!), the stirrer
switches on simultaneously.