Minifors 2 - Benchtop Bioreactor
Technical Data
07 February 2019
Page
213
of
214
14.5.9 Pumps
Type
Quantity
Peristaltic
4
Configuration
Standard
Alternative setting
Pump 1
ACID,
digital
FEED,
(analogue)
Pump 2
Base,
(digital)
FEED,
(analogue)
Pump 3
AF
= (antifoam), digi-
tal
LEVEL, (digital)
or:
FEED, (analogue)
Pump 4
FEED,
(analogue)
BALANCED, (ana-
logue)
Hoses
Standard
Option 1
Option 2
Inside diameter 1.0 mm
0.5 mm
2.5 mm
Delivery rate
1
3.5 ml min
-1
1.1 ml min
-1
16.1 ml min
-1
Material
PharMed BPT
1
)
Typical figure with water measured at max. rotation speed
14.6 Operating Conditions
Description
Value
Temperature range
5 °C up to 40 °C
Relative air humidity, non-con-
densing
20 % up to 90 %
Min. distance from walls, ceil-
ings and other appliances
150 mm
14.7 Emissions
Description
Value
Units
Noise emission
<70