3-46
ES User’s Guide
Making Measurements
Measuring S21 Forward Transmission using an Enhanced Response Calibration
View and Interpret the S21 Measurement
Results
1. To view the entire measurement trace on the display, press
.
2. To interpret the measurement, refer to
or your analyzer's display if you are
making this measurement on your instrument.
a. The values shown on the horizontal axis are the frequencies in
MHz. The values shown on the vertical axis are the power ratios
in decibels (dB) of the transmitted signal through the device
divided by the incident power. To display the result in logarithmic
magnitude format (designated by "Log Mag" at the top of the
measurement screen), the analyzer computes the measurement
trace using the following formula:
where P
trans
= the power transmitted through the device and
where P
inc
= the incident power.
b. A level of 0 dB would indicate a perfect through cable or device (no
loss or gain). Values greater than 0 dB indicate that the DUT has
gain. Values less than 0 dB indicate loss.
3. To quickly determine the filter's minimum insertion loss, press
.
4. Note the marker readout in
amplitude of the minimum insertion-loss point.
SCALE
Autoscale
Transmission dB
(
)
10 log
P
trans
P
inc
--------------
=
MARKER
Marker Search Max Search Mkr
−
> Max
Summary of Contents for HP 8712ES
Page 8: ...viii ES User s Guide ...
Page 17: ...ES User s Guide 1 1 1 Installing the Analyzer ...
Page 34: ...1 18 ES User s Guide Installing the Analyzer Preventive Maintenance ...
Page 35: ...ES User s Guide 2 1 2 Getting Started ...
Page 52: ...2 18 ES User s Guide Getting Started Performing the Operator s Check ...
Page 53: ...ES User s Guide 3 1 3 Making Measurements ...
Page 135: ...ES User s Guide 4 1 4 Using Instrument Functions ...
Page 159: ...ES User s Guide 4 25 Using Instrument Functions Using Markers Figure 4 15 Delta Marker Mode ...
Page 165: ...ES User s Guide 4 31 Using Instrument Functions Using Limit Testing Figure 4 16 Limit Lines ...
Page 224: ...4 90 ES User s Guide Using Instrument Functions Using an External VGA Monitor ...
Page 225: ...ES User s Guide 5 1 5 Optimizing Measurements ...
Page 246: ...5 22 ES User s Guide Optimizing Measurements Measuring Devices with Long Electrical Delay ...
Page 247: ...ES User s Guide 6 1 6 Calibrating for Increased Measurement Accuracy ...
Page 281: ...ES User s Guide 7 1 7 Front Rear Panel ...
Page 306: ...7 26 ES User s Guide Front Rear Panel Line Module ...
Page 307: ...ES User s Guide 8 1 8 Hardkey Softkey Reference ...
Page 410: ...8 104 ES User s Guide Hardkey Softkey Reference Z ...
Page 411: ...ES User s Guide 9 1 9 Specifications ...
Page 469: ...ES User s Guide 10 1 10 Safety and Regulatory Information ...
Page 475: ...ES User s Guide 10 7 Safety and Regulatory Information Regulatory Information ...
Page 476: ...10 8 ES User s Guide Safety and Regulatory Information Regulatory Information ...
Page 477: ...ES User s Guide 11 1 11 Factory Preset State and Memory Allocation ...